HACH LXV322 TSS Portable Turbidity, Suspended Solids, and Sludge Level System

Brands:Hach

Dijual & dikirim oleh: Kucari.com
  • Resolution
  • Turbidity: 0.001 at 0 to 0.999 NTU; 0.01 at 1 to 9.99 NTU;
  • 0.1 at 10 to 99.9 NTU; 1 at greater than 100 NTU
  • Suspended Solids: 0.001 at 0 to 0.999 g/ L; 0.01 at
  • 1 to 9.99 g/ L; 0.1 at 10 to 99.9 g/ L; 1 at greater than 100 g/ L

HACH LXV322 TSS Portable Turbidity, Suspended Solids, and Sludge Level System

Description :

System
OverviewDetailsParameter/ReagentDownloadsAccessoriesSimilar ProductsTSS Portable Hand-held Turbidity, Suspended Solids, and Sludge Level System Product #:LXV322.99.00002USD Price:Contact Hach Available

Three Parameters in One Instrument
The Hach TSS Portable Hand-held Turbidity, Suspended Solids, and Sludge Blanket Level Instrument’s unique multi-beam alternating light method with infrared diode system gives it a broad measuring range for both turbidity and suspended solids in one portable handheld instrument.

Multiple Calibration Curves for Convenience
It stores up to four different calibration curves for suspended solids and one for turbidity. There’s no need for constant re-calibration for different locations or types of sludge.

Easy Sludge Blanket Levels
The sludge blanket level can be measured by setting the units to mg/L, g/L or percent. Calibrate the instrument and then slowly lower the sensor into the clarifier. The suspended solids concentration will significantly increase as it reaches the sludge blanket level. Once the blanket level is reached, examine the sensor cable which is marked every meter to determine the sludge blanket level depth from the water surface.

Air Bubble Compensation for Accuracy
The instrument software automatically compensates for air bubbles to prevent its influence on the measurement. Turbidity is pre-calibrated with a calibration curve stored in the instrument. For suspended solids, each of the calibration curves can be closely adjusted to the laboratory results because each calibration curve can use up to three points per curve plus a zero value.

Durable Materials for Long Life
The probe is made of polished stainless steel with a scratch-resistant sapphire window to withstand harsh environments. Polished stainless steel helps keep particles from sticking to the surface of the probe. The internal memory stores up to 290 measuring values, including time, date, measurement, location, and homogeneity. Three operating modes are available: single, interval, or continuous.

Specifications :

Measurment Principle
Combined multi-beam alternating light method with infrared
diode system      IW
Turbidity: 2-channel 90ï ¿ ½ scattered light measurement in
accordance with DIN EN 27027/ ISO 7027; additional
measured value verification by six-channel multiple angle     E
measurement
Suspended Solids: modified absorbance measurement, six-channel multiple angle measurement

Range
Turbidity: 0.001 to 4000 NTU
Suspended Solids: 0.001 to 400 g/ L ( 1 to 400, 000 mg/ L) ( upper range depending on sample characteristics)

Accuracy
Turbidity: less than 3% or ï ¿ ½ 0.02 NTU, whichever is greater Suspended Solids: less than 4% or ï ¿ ½ 0.001 g/ L whichever is greater

Resolution
Turbidity: 0.001 at 0 to 0.999 NTU; 0.01 at 1 to 9.99 NTU;
0.1 at 10 to 99.9 NTU; 1 at greater than 100 NTU
Suspended Solids: 0.001 at 0 to 0.999 g/ L; 0.01 at
1 to 9.99 g/ L; 0.1 at 10 to 99.9 g/ L; 1 at greater than 100 g/ L

Units of Measure
Turbidity: NTU, FNU, EBC
Suspended Solids: ppm, mg/ L, g/ L, %
Operating Modes
Single, interval, or continuous measurement

Display
24 mm ( 0.95 in.) high LCD graphic display; UV treated; alphanumeric 4 lines of 16 characters each
Operating Temperature
0 to 60ï ¿ ½ C ( 32 to 140ï ¿ ½ F)
Operating Pressure
0 to 10 bar ( 0 to 145 psi)
Relative Humidity
0 to 95% relative humidity
Power Requirements
Meter: 6 rechargeable NiMH batteries
( recommended 1.2 V/ min. 1800 mAH) or
6 ï ¿ ½ AAï ¿ ½ standard batteries
Plug-in Charger: 115/ 230 Vac, 50/ 60 Hz
Sensor Cable Length
10 m ( 32.8 ft.)
Meter Enclosure Rating
IP55
Probe Enclosure Rating
IP68
Current Input
Approx. 60 mA
Wetted Materials Probe: stainless steel Sensor Window: sapphire
Internal Data Memory
For up to 290 measured values
Weight
Meter: 0.6 kg ( 1.2 lb.)
Probe: 1.6 kg ( 3.5 lb.)
Dimensions
Meter: 110 x 230 x 40 mm
( 4.3 x 9.1 x 1.6 in.)
Probe: 290 x 40 mm ( 11.4 x 1.6 in.)
Case: 320 x 450 x 110 mm
( 12.6 x 17.7 x 4.3 in.)

HACH LXV322 TSS Portable Turbidity, Suspended Solids, and Sludge Level System Dengan Harga murah dan cari produk terbaik lainnya dari kategori  dari Turbidity Meter  Merek Hach Gratis biaya pengiriman untuk wilayah Jakarta dan dapat dikirim ke seluruh Indonesia. Dapatkan hanya di toko online Kucari.com, belanja online alat ukur dan alat survey mudah dan aman.

Kucari.com menyediakan berbagai alat ukur dan instruments dari supplier dan distributor resmi yang tersebar dibeberapa kota di Indonesia seperti Jakarta, Bandung, Bekasi, Depok, Cikarang, Tangerang, Lampung, Pontianak, Jogja, Semarang, Solo, Batam, Padang, Pekanbaru, Malang, Surabaya, Medan, Makassar, Balikpapan, Samarinda, Pelembang, Riau, Aceh, Manado, dan Bali.

Daftarkan perusahaan Anda!

Anda adalah Agen, Vendor, Supplier & Distributor peralatan ukur dan perkakas untuk wilayah Indonesia? Daftarkan perusahaan Anda disini

Kucari.com merupakan situs penyedia direktori peralatan dan perkakas bisnis sebagai media promosi online berbasis web yang dapat menghubungkan supplier, distributor, agen, vendor dan pembeli antar bisnis (B2B – Business to Business) dan kalangan B2C (Business-to-Customer) dalam 24 Jam nonstop menjangkau seluruh wilayah Indonesia

Based on 0 reviews

0.0 overall
0
0
0
0
0

Be the first to review “HACH LXV322 TSS Portable Turbidity, Suspended Solids, and Sludge Level System”

There are no reviews yet.